- This is typical of when the instrument is running in the "regulated" flow rate. This mode is for when there is insufficient CFM (air flow rate) from the compressor to run the LM10 continuously, the pause gives the compressor time to catch up. Check the pages on the touch screen and verify the running mode.
- If it is in "regulated" then switch to "free" mode.
- If it is still in free mode, we may need to remote diagnose further and possibly come to inspect.